Parts & controls – Amprobe AMB-35 Insulation-Tester User Manual

Page 3

Advertising
background image

3

4

‰

DC Voltage

c

LCD display.

Range Resolution Accuracy Input Impedance Overload Protection
1000V

-1000V

1V

1%+1

2.5MΩ

1000Vrms

d

MEM key: Manual data memory control key.

e

LOCK key: Locks the Insulation test Resistance (MΩ) and Low resistance

(LoΩ) functions.

Sensitivity : 2V

a). Press and hold down TEST key, then press LOCK key, the

“ LOCK ” icon appears on the display.

‰

Resistance & Continuity

Range

Resolution

Accuracy

Max. open Circuit

Voltage

Overload

Protection

9999Ω

1%+2

3V

1000Vrms

Touch the probe tips together, then press

key,

the

icon appears on the display and the main

display indicates 0.00Ω.

b). In MΩ mode continuously applies the test voltage to the circuit to

be tested. The beeper sounds every 2 seconds to remind you
that you are in the LOCK mode.

c). In LoΩ mode continuously applied the test current to the circuit to

be tested.

3. PARTS & CONTROLS

READ

MEM

s

AC

m:

HOLD

VG

K

M

COM

V

d). Press Lock or TEST key again to exit the mode.

f

READ key: Manual memory data reading control key.

g

key: In the Lo Ω function, turns the test lead resistance

compensation ON.

h

TEST key:

a). Used for MΩ and LoΩ test functions.
b). Press and hold TEST key until the main display reading is stable.

i

j

▼▲ keys : In READ mode, select the memory data location for

direct reading the data from the display.

k

Rotary switch: To select a measurement function.

l

Input terminals.