I N T E R F AC E M O D E L 9 8 4 0 V E R 6 . 0 . 2

#

Message

Meaning

1

Interface X.Y.Z

This is the version number.

2

Serial # 12345

Option # 114236

This is the serial number.

This is the option number.

3

Last ID A LD90437

Last ID B TQ12030

The serial number of the load cell on Ch A.

The serial number of the torque cell on Ch B

4

Ch A Max 4.00Lb

Ch B Max 4.00LbI

Maximum rated load of the sensor on channel A.

Maximum rated torque of the sensor on channel B.

5

Ch A Cal 4.00000 mVv

Ch B Cal 4.00000 mVv

The mV/V constant used with the load cell on channel A.

The mV/V constant used with the torque cell on channel
B.

Note: If either of these sensors was calibrated by 5 –Point
Mass, then 4 mV/V constants will be displayed here.

6

Ch A Exc 5.00 Volt

Ch B Exc 10.00 Volt

Excitation voltage used with the load cell on channel A.

Excitation voltage used with the torque cell on channel B.

Each of these messages will be displayed for about 3 seconds. After that the front display
will start showing the current readings from the sensor.

MODEL 9840 FRONT PANEL


The example above shows that Load A is 0.0000 Lb. The second line of the display
shows text.

To change what item is being displayed press the leftmost button. This will cycle through
the list of items (load A, peak A, valley A, gross A, load B, peak B, valley B, gross B,
channel A+B, pos, velocity, lim). If a torque cell is loaded on either channel, the display
will show torq rather than load for the appropriate channel. Channel A+B will only appear
when both channels are attached to sensors of the same type: load or torque.