HP 2040D User Manual

Page 47

Advertising
background image

Series 2040 Test System

Series 2040 Maintenance Manual V2.00

Certification

46

Calibration Station. Finally, connect the GPIB port from the 2040 console to

the GPIB port on the Calibration Station.

To run the program select the

certification icon from the Atart/

Programs/Digalog program group,

and the screen to the right will

appear prompting for the type of

calibration. Select the TDAC, and

the screens in the right column on

both this page and the following

page will appear in the order

shown.

The first screen prompts for the

user’s name, the serial number of

the 8840 Fluke multimeter in the

Calibration Station. This number can be found on a Digalog tag inside the

Calibration Station next to the

meter. It also prompts for the

Selftest Assembly serial number,

and the serial number of the

tester itself.

When the information is

entered, the last setup screen

will appear requesting a

verification of the information

that was entered. If any of the

information is incorrect, it can

be modified or changed at this

time.

The operator’s name, Selftest

serial number, and Fluke 8840A

meter serial number must be

entered and verified before the

unit can be certified. Click on the “Certify” button and the program checks

the TDAC signal. Test results and a pass/fail condition are displayed in the