Advanced operations – Hach sc1000 User Manual

Page 87

Advertising
background image

85

Advanced Operations

Relation between input current and calculated concentration

Figure 52

shows the output value depending on the input current, the set low value and

the set high value with an input range of 0

20 mA.

The output value (OV) is a function of the input current (IC).

The output value is determined according to formula

(3)

:

(3)

where:
OV=Output value
IC=Input current
LV=Low value
HV=High value

Figure 52 Output value with an input range of 0

20 mA

1

Output value (concentration) (x-axis)

5

0 mA

2

OV=f(IC)

6

Low value (LV)

3

Input current (IC) (y-axis)

7

High value (HV)

4

20 mA

OV

f(IC)

IC

HV-LV

20mA

-----------------

×

LV

+

=

=