Table 3, Ous location, Installation – Hach 6120118 User Manual

Page 13

Advertising
background image

13

Installation

3.2 Connecting/Wiring the Digital Gateway to the sc100 Controller

DANGER
The sc100 and certain versions of the sensor are suitable for use in Class 1,
Division 2, Groups A, B, C, D Hazardous Locations . See Control Drawing 58600-78
in the sc100 Controller Manual, Cat. No. 58600-18 for acceptable sensor versions
and installation requirements.

DANGER
Le sc100 et certaines versions du capteur peuvent être utilisés dans des endroits
dangereux de la Classe 1, Division 2, Groupes A, B, C, D. Reportez-vous au schéma
de contrôle 58600-78 du Manuel du contrôleur sc100, Réf. 58600-18 pour connaître
les versions des capteurs admises et les conditions d'installation.

The digital gateway should be wired to the sensor before connecting to the controller.

3.2.1 Connecting the Digital Gateway Sensor in a Non-hazardous Location

3.2.1.1 Attaching the Digital Gateway with a Quick-connect Fitting in a Non-hazardous Location

Important Note: The standard quick-connect fitting is NOT suitable for Class 1, Division 2
Hazardous Location installations without the connector lock installed, see

section 3.2.1 on

page 13

for more information.

The digital gateway is supplied with a keyed quick-connect fitting for easy attachment
to the controller, see

Figure 7

. Retain the connector cap to seal the connector opening in

case the sensor must be removed. Optional extension cables may be purchased to extend
the sensor cable length. If the total cable length exceeds 100 m (328 ft), a termination box
must be installed. See

Replacement Parts and Accessories on page 37

.

Table 3 Wiring the Digital Gateway (Cat. No. 6120800)

Sensor (wire color)

Sensor Signal without Ground Rod

Sensor Signal with Ground Rod

1

Digital Gateway

Metal Braid

2

Jumper 2 from J3-1 to J3-3

3

Ref

J3-1

White

Temp +

Temp +

J3-2

Red

Temp –

Temp –

J3-3

Blue

Ref

Solution Ground

J3-4

Clear

Active/Measuring

Active/Measuring

J1-5

not used

not used

not used

J1-6

1

Some applications require the use of an external ground rod with the combination electrode. Use this wiring scheme for
these applications.

2

If Metal Braid is “tinned”, cut the tinned area off and twist the wire to insert into connector.

3

Customer-supplied (required)