Advanced Operations

7. The result is displayed and is to be compared with the

nominal/standard data of the test filter. If a specific sample is
used, the measured results must be evaluated against
user-defined requirements.

8. Press Cancel to return to Optical Checks.

6.8.3.5 Drift check

The Drift Check test is used to test the stability of the instrument.

Note: The Drift Check runs 1 hour.

1. Press Drift Check in the Optical Checks menu.

2. Press

λ to input the wavelength.

3. Input the wavelength and press OK to confirm.