Advanced operations – Hach DR 5000 User Manual

Page 114

Advertising
background image

114

Advanced Operations

3. Press Edit.

An automatic menu guidance queries values (filters,
wavelength, nominal values and tolerances) given in the quality
control certificate, to the following specifications:

Stray Light

Photometrical accuracy

Wavelength accuracy

4. Press OK when all values are entered and the overview is

displayed.

5. Press Verification.

6. Insert the Multi Cell Holder in the cell compartment in such a

way that the cell opening for 10 mm square cuvettes/cells is at
the front. Secure the holder with two locking screws.

7. Remove any cuvettes/cells from the cell compartment, close

the cell compartment, and press Start.