Operation, Immersion depth and usage – Fluke STIK 1551A EX User Manual

Page 13

Advertising
background image

Stik Thermometer

Operation

5

Operation

Immersion Depth and Usage

W

Caution

To prevent possible damage to the Product or to equipment

under test, do not immerse the Probe beyond the maximum

immersion depth. This may damage its internal electronics.

Allow enough time for the Probe to stabilize before recording measurements. Use the

Stability Display to determine when the Probe has stabilized.
Probe sensor length is shown in Figure 1.

3

2

1

gjo004.eps

A

Sheath Length:
1551A-9: 229 mm (9 in)
1551A-20: 508 mm (20 in)
1551A-12/1552A-12: 305 mm (12 in)

B

Sensor Length:
1551A:

10 mm (0.39 in)

1552A:

30 mm (1.18 in)

C

Sheath Diameter:
1551A-9: 4.8 mm (3/16 in)
1551A-12/1551A-20/1552A: 6.35 mm (1/4 in)

Figure 1. Probe Measurements