1551A/1552A Ex

Users Manual

6

The minimum immersion depth for the Probe is 3.5 inches in a stirred fluid heat source. It

is recommended that you make sure of the minimum immersion depth for the specific

application. See Figure 2.

The maximum immersion depth for the 1551A Ex is 2 inches less than the total

length. Make sure the last 2 inches of Probe, nearest to the internal electronics of the

device, are not put into the heat source.

The maximum immersion depth for the 1552A Ex is 3 inches less than the total

length. Make sure the last 3 inches of the Probe, nearest to the internal electronics of

the device, are not put into the heat source.

gjo002.eps

A

1552A Ex: Maximum immersion: 3 inches less than total probe length.

B

1551A Ex: Maximum immersion: 2 inches less than total probe length.

Figure 2. Immersion Depth