To estimate capacitance values above 5

µ

F, use the

current supplied by the meter’s resistance function,
as follows:

1. Set up the meter to measure resistance.

2. Press

K

to select a range based on the

value of capacitance you expect to measure
(refer to Table 6.)

3. Discharge the capacitor.

4. Place the meter’s leads across the capacitor;

then time how long it takes for the display to
reach OL.

5. Multiply the charge time from step 4 by the

appropriate value in the

µ

F/second of Charge

Time column in 6. The result is the estimated
capacitance value in microfarads (

µ

F).

Table 6. Estimating Capacitance Values Over

5 Microfarads

Expected Capacitance

Suggested

Range*

µ

F/second

of Charge

Time

Up to 10

µ

F

4 M

e

0.3

11

µ

F to 100

µ

F

400 k

e

3

101

µ

F to 1000

µ

F

40 k

e

30

1001

µ

F to 10,000

µ

F

4 k

e

300

10,000

µ

F to 100,000

µ

F

400

e

3000

*

These ranges keep the full-charge time between 3.7 seconds

and 33.3 seconds for the expected capacitance values. If the
capacitor charges too quickly for you to time, select the next
higher resistance range.