Table 7. Functions and Trigger Levels for Frequency Measurements

Function

Range

Approximate

Trigger Level

Typical Application

K

4 V, 40 V, 400 V,
1000 V

0 V

Most signals.

K

400 mV

0 V

High-frequency 5 V logic signals. (The dc-coupling of the

L

function can

attenuate high-frequency logic signals, reducing their amplitude enough
to interfere with triggering.)

L

400 mV

40 mV

Refer to the measurement tips given before this table.

L

4 V

1.7 V

5 V logic signals (TTL).

L

40 V

4 V

Automotive switching signals.

L

400 V

40 V

Refer to the measurement tips given before this table.

L

1000 V

400 V

ReEG

Frequency counter characteristics are not specified for these functions.

\

All ranges

0 A

AC current signals.

µ

A

F

400

µ

A

Refer to the measurement tips given before this table.

^

40 mA

A

F

4 A