Emissivity - Table and Unknown Value

To adjust the unit’

TT

s

emissivity value for a

material with unknown

emissivity, plug in the

probe.

Pull the unit’s trigger.

Place the measuring

tip of the probe on the

area to be measured.

Wait for the reading to

stabilize.

P

7

5

C

D

Note the indicated probe temperature reading.

Release the trigger. Pull the trigger again. Measure

the same area using infrared measurement. Press

the emissivity button (P). Use the Up and Down keys

(C) to select the material name “Free” which will be

shown in the display (7). Press the emissivity button

(P) again until the emissivity sign (5) flashes. Use

the arrow keys (C) to change the emissivity value

until the temperature matches the probe’s reading.

UNKNOWN VALUE

To choose the emissivity

of a material, press

EMISS (P). The display

shows a material name

(7), an emissivity value,

and the calculated

temperature value (5). To

choose another material,

use the Up and Down

keys (C). Press ENTER

(D) to activate this setting.

TABLE OF VALUES