When the meter is set to measure frequency and there is no input signal
(i.e., the input terminals are open), the meter may read approximately 25
kHz (rather than zero). This is due to internal capacitive pickup of the
inverter power supply into the high-impedance, input circuitry. With source
impedance of <2 k

, this pickup will not affect the accuracy or stability of

the frequency reading.

Environmental

Warmup time

Temperature Coefficient

Operating Temperature

Storage Temperature

Relative Humidity
(non-condensing)

Altitude

Electromagnetic
Compatibility

Vibration

Shock

1 hour to rated specifications for warmup < 1 hour, add 0.005 % to all
accuracy specifications.

<0.1 times the applicable accuracy specification per degree C for 0 °C to
18 °C and 28 °C to 50 °C (32 °F to 64.4 °F and 82.4 °F to 122 °F)

0 °C to 50 °C (32 °F to 122°F)

-40 °C to + 70 °C (-40 °F to 158°F)

Elevated temperature storage of battery will accelerate battery
self-discharge. Maximum storage time before battery must be
recharged:

20 °C – 25 °C

1000 days

50 °C

180 days

70 °C

40 days

To 90 % at 0 °C to 28 °C (32-82.4 °F),

To 80 % at 28 °C to 35 °C (82.4-95 °F),

To 70 % at 35 ° C to 50 °C (95 °F -122 ° F) except to 70 % at 0 °C to 50
°C (32 °F -122 °F) for the1000 k

, 3 M

, 10 M

, 30 M

, 100 M

,and

300 M

ranges.

Operating

0 to 10,000 feet

Non-operating

0 to 40,000 feet

In an RF field of 1 V/m on all ranges and functions: Total Accuracy =
Specified Accuracy +0.4% of range. Performance above 1 V/m is not
specified

3 G @ 55 Hz

Half sine 40 G. Per Mil-T- 28800D, Class 3, Style E.

Bench Handling. Per Mil-T-28800D, Class 3.