General – Fluke 45 User Manual

Page 114

Advertising
background image

45

Users Manual

General

Common Mode Voltage

Size

Weight

Power

Standards

RS-232-C

1000 V dc or peak ac maximum from any input to earth

9.3 cm high, 21.6 cm wide, 28.6 cm deep (3.67 in high,8.5 in wide,
11.27 in deep)

Net, 2.4 kg (5.2 Ibs) without battery;

3.2 kg (7.0 Ibs) with battery;

Shipping, 4.0 kg (8.7 Ibs) without battery;

4.8 (10.5 Ibs) with battery

90 V to 264 V ac (no switching required), 50 Hz and 60 Hz < 15 VA
maximum

Complies with: IEC 348, UL1244, CSA Bulletin 566B

EMC: Part 15 subpart J of FCC Rules, and VDE 0871.

Baud rates: 300, 600,1200,2400,4800 and 9600

Odd, even or no parity

One stop bit

Options

Battery (Option -01 K

)

IEEE-488 (Option

-

05K

)

Type

8 V, Lead-Acid

Operating Time

8 hours (typical).

N

lights when less than

1/2 hour of battery operation remains.
Meter still meets specifications.

Recharge Time

16 hours (typical) with meter turned off
and plugged into line power. Battery will
not charge when meter is turned on.

Capability codes

SH1, AH1, T5, L4, SRI, RL1, PP0, DC1,

DT1, E1, TED, LEO and C0

External Trigger Input

VIH

1.35 V minimum

VIL

1.25 V maximum

Input Threshold Hysteresis

0.6 V minimum

Advertising