Operating the Meter From the Front Panel

Selecting A Function Modifier

3

3-11

CAL

ENABLE

V

10A

100

mA

600V CAT I

1000V CAT I

FUSE F1

500 mA

F 250V

POWER

AUTO

2ND

MN MX

RATE

dB

REL

REF#

LOCAL

THRESH

ADDR

BAUD

HOLD

LO

HI

COMP

A

A

V

FREQ

DUAL DISPLAY MULTIMETER

45

V

!

REF

FUSED

COM

MN MX

dB

REL

HOLD

-Press to Toggle In and Out of Relative Modifier

-Press to Toggle In and Out of Decibels Modifier

-Press to Select Touch Hold Modifier;
-Press to Force New Reading;
-Press and Hold Down for
2 Seconds to Exit.

-Press to Select MN MX Modifier;
-Press to Toggle Between Minimum
and Maximun Reading;
-Press and Hold Down for 2
Seconds to Exit MN MX Modifier.

aam15f.eps

Figure 3-7. Function Modifier Selection Buttons

REL (Relative Readings) Modifier

When the relative modifier (REL) is selected, the reading on the primary display is
always the difference between the relative base and an input measurement. For example,
if the relative base is 15.000 V, and the present reading is 14.100 V, the display will show
-0.900.

W

Warning

To avoid electrical shock or damage to the meter, note that a
relative reading may not indicate the presence of dangerous
voltages at the input connectors or test leads.

Press

K to toggle in and out of the relative modifier. When the relative modifier is

selected: the last valid reading is stored as the relative base, the primary display zeroes
out, and "

K" is shown on the primary display. (The secondary display is unaffected.)

To edit the relative base, use the number editor as described in "Using the Number
Editor" later in Chapter 3).

Note

The relative modifier cannot be selected I the display shows "OL" or is
blank. (The display would be blank, for example, because of external
triggering or range changes.)

Selecting the relative modifier

K turns off autoranging and locks in the present range.

Make sure you are in the correct range before selecting the relative modifier. If you press

U or T after the relative modifier has been selected, you will automatically exit K.
When you are in REL, the relative base can be shown in the secondary display by
pressing

S and, while holding it down, pressing K. Now, whenever the relative

Advertising