Troubleshooting – Fluke Thermometer 576 User Manual

Page 28

Advertising
background image

28

576

Troubleshooting

Symptom

Display Code

Problem

Action

Camera cannot

Driver installation

Open

send photos

failed

Harware Manager on

to the PC

the PC to fix it

-O-

Target temperature is over

Select target

-U-

or under range

within unit’s specs

EEPROM-Err

EEPROM error

Contact Factory

CalAreaErr

calibration errors

Contact Factory

ProbCalEr

Flash needs

Battery is low

Replace Batteries

too long to
charge or does not
charge at all

Blank display

Battery is dead

Replace Batteries

Laser won’t

Low or dead battery

Replace Batteries

work

Ambient above 45°C (113°F)

Operate unit in
45°C (113°F)
ambient or below

Display “ON”

Display locked “ON”

Disconnect the unit

from the PC

Photos lost

Batteries changed before

Wait for shutdown

shutdown or too long

to end. (blank screen)

after removing old ones

Change batteries within

two minutes of

removing old ones

PC displays “logger The setup in the unit and

Click “OK”. Either save

file does not match

in the PC software do not

the current data to a

device logger setup” match

new file name or find

the correct *.lgg file,

open it in “Setup

Device Logger”, save,

and redownload the

data with the “Data

Logger/Load Device

Data Logger”

Sometimes a long beep and the red LED above
the display shines to indicate a problem, check
your display to find out which problem has
occured.