Calibration adjustment, Test equipment – Fluke 700G User Manual

Page 20

Advertising
background image

700G Series

Calibration Manual

10

Table 3. Verification Points (In PSI) (cont.)

700G31 10000 PSI

Input Pressure

Lower Limit

Upper Limit

10000 9995 10005

9000 8995 9005

7500 7495 7505

6500 6495 6505

5500 5495 5505

4500 4495 4505

1500 1495 1505

500 495 505

0 -5 5

*At altitudes -14 psi or -12 psi may not be possible. Replace with a local near-vacuum point when
necessary.

Calibration Adjustment

Calibration adjustment is done electronically by internal software with the Product case

closed. All calibration commands and adjustments are done with the keypad.
Eight calibration points are used in the adjustment. They go from full scale to zero at

pressures that equal 100 %, 87.5 %, 75 %, 62.5 %, 50 %, 37.5 %, 25 %, 12.5 %, and 0 %

of full scale plus vacuum.

Note

This is an ambient temperature calibration, and must be done at an ambient
temperature of 23 °C ± 3 °C (72 °F ± 5 °F). Calibration out of this
temperature range voids the temperature compensation software in the
Product.

Test Equipment

For verification and calibration adjustment, pressure and/or vacuum standards that can

make and show pressures from vacuum to the full-scale range of the UUT are necessary.

A TUR standard of 4:1 or better is necessary to keep the Product at its specified accuracy.