700G Series

Calibration Manual

12

Note

Only some ranges use vacuum calibration. If your Product does not,
then this step can be skipped and calibration is complete.

10. Use the Pressure Standard to output -12.00 psi. After the output is stable, push  to

continue. As the Product measures, the screen will show ———-. When the

measurements are complete, a Product reset occurs and the Product turns off then

turns on as usual.

Table 4. Calibration Points

Model

Number

700G04

700G05 700G06 700G27

700G07

700G08

700G29

700G30

700G31

Pressure Range (psi)

15

30

100

300

500

1000

3000

5000

10000

100

%

15.00

30.00 100.00 300.00 500.00 1000.0 3000.0 5000.0 10000

87.5

%

13.125 26.25 87.50 262.50 437.50 875.00 2625.0 4375.0 8750.0

75

%

11.25

22.50 75.00 225.00 375.00 750.00 2250.0 3750.0 7500.0

62.5

%

9.375

18.75 62.50 187.50 312.5 625.00 1875.0 3125.0 6250.0

50

%

7.50

15.00 50.00 150.00 250.00 500.00 1500.0 2500.0 5000.0

37.5

%

5.625

11.25 37.50 112.50 187.50 375.00 1125.0 1875.0 3750.0

25

%

3.75

7.50 25.00 75.00 125.00 250.00 750.00 1250.0 2500.0

12.5

%

1.875

3.75 12.50 37.50 62.50 125.00 375.00 625.00 1250.0

0

%

0.00

0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00

Serial Interface

Use terminal communication software on a PC to set up terminal communication. An RS-

232 to USB cable is necessary. This cable comes with 700G/TRACK Software. Use these

terminal parameters:
• Bits per second: 9600
• Data bits: 8
• Parity: None
• Stop bits: 1
• Flow control: None
• Local echo: on