2 field settings (f2), Table 15 default statistical fields, Table 16 statistical field descriptions – Fluke 1595A User Manual

Page 58: Field settings (f2)

Advertising
background image

1594A/1595A Super-Thermometer

Main Menu

48

When the User Settings screen appears after power-on, when the Start-Up Help setting is ON, the User Set-

tings screen includes function key NEXT (F1). This key leads to the Time and Date Settings screen (see Sec-

tion 5.7.5.1, TIME DATE (F1), on page 50), where the clock may be set to the local time.

5.7.4.2 FIELD SETTINGS (F2)

MAIN MENU | DISPLAY MENU |

The Field Settings function key is used for selecting statistical fields to display in the Measurement Screen. A

particular set of statistical fields can be stored as a set and recalled using the RECALL MEMORY selection.

Four different sets can be stored in Recall Memory, any of which can be restored to factory defaults using the

RESTORE DEFAULT (F1) function key (see Default Statistical Fields table below). Restore Default only af-

fects the Recall Memory selection that is currently display.
Individual statistical fields in any of the Recall Memory sets can be changed by first selecting the RECALL

MEMORY set number, pressing the ENTER key, arrowing down to the field to be changed, changing the field

with the Left/Right arrow keys and pressing the ENTER key save the change. The available data field options

and their descriptions are listed in the tables below.

Table 15 Default Statistical Fields

Field

Memory 1

Memory 2

Memory 3

Memory 4

1

MEAN

MEAN

MAXIMUM

MEAN

2

STD DEV

STD DEV

MINIMUM

STD DEV

3

SE OF MEAN

N

DELTA

TIME

4

RESISTANCE

RESISTANCE

RESISTANCE

DATE

Table 16 Statistical Field Descriptions

Option

Description

Equation

MEAN

arithmetic mean of measurements

Mean

n

y

i

n

i

=

=

1

1

STD DEV

sample standard deviation of measurements

Std Dev

n

y

n Mean

i

i

n

=

− •







=

1

1

2

1

2

SE OF MEAN

standard error (estimated standard deviation) of the
mean

SE of Mean

std dev

n

= (

)

MAXIMUM

maximum measurement

MINIMUM

minimum measurement

SPREAD

difference between maximum and minimum

Spread Maximum Minimum

=

N

number of measurements (sample size)

DELTA

difference relative to a delta channel or offset

RESISTANCE

resistance measurement

RATIO

resistance ratio measurement

TIME

time from the system clock

DATE

date from the system clock

As a note of further explanation, SE OF MEAN provides an estimate of the uncertainty (k = 1) of the MEAN

value. SE OF MEAN is calculated from the sample standard deviation of unfiltered measurements (not the

same as the displayed standard deviation of filtered measurements) and the sample size. The method used to

calculate SE OF MEAN disregards the effect of the digital filter since it causes correlation among displayed

measurements. Otherwise the result of the formula for SE OF MEAN would produce an erroneously low

value.