7 *stb, 8 cal:date <year>,<month>,<day, 9 cal:par:lin<n> <value – Fluke 1595A User Manual

Page 86: 10 cal:par:res<n> <value, 11 cal:par:zero <value, 7 *stb? 8.7.3.8, Cal:date <year>,<month>,<day, Cal:par:lin<n> <value, Cal:par:res<n> <value, Cal:par:zero <value

Advertising
background image

1594A/1595A Super-Thermometer

Interface Commands

76

Query form: *SRE?
Sets the Service Request Enable register. This mask register determines which corresponding bits of the Status

Byte Register (see Section 8.7.3.7, *STB?, on page 76) affect the MSS master summary bit (bit 6) of the

Status Byte Register and generate a service request from the IEEE-488 port.

8.7.3.7 *STB?

Example response: 16
Reads the Status Byte Register. This indicates any of several events defined by the IEEE-488.2 and SCPI stan-

dards and which are listed in the following table. Reading the Status Byte Register does not affect it.

Bit

Condition

0

(not used)

1

(not used)

2

Error

3

Questionable status

4

Message available

5

Event status

6

Master summary status

7

Operation status

8.7.3.8 CAL:DATE <year>,<month>,<day>

Example command: CAL:DATE 2009,8,21
Query form: CAL:DATE?
Password protection: unconditional
Sets the instrument’s date of calibration.

8.7.3.9 CAL:PAR:LIN<n> <value>

Query form: CAL:PAR:LIN1?
Password protection: unconditional
Sets a calibration Linearity C<n> parameter.
These settings may be used to apply a linearity correction to the measurement circuit.

IMPORTANT:

Calibration parameters should not be changed except as part of the Super-Thermometer

calibration procedure performed by a qualified service technician.

8.7.3.10 CAL:PAR:RES<n> <value>

Query form: CAL:PAR:RES<n>?
Password protection: unconditional
Sets the calibrated resistance of an internal reference resistor. Suffix <n> specifies the resistor; 1: 1 Ω; 2: 10 Ω;

3: 25 Ω; 4: 100 Ω; 5: 10 kΩ.
IMPORTANT: Calibration parameters should not be changed except as part of the Super-Thermometer cali-

bration procedure performed by a qualified service technician.

8.7.3.11 CAL:PAR:ZERO <value>

Query form: CAL:PAR:ZERO?
Password protection: unconditional