DOS 3 (neutralizer)

Setting the neutralizer dosage

Set dosage concentration according to the neutralizer manufacturer's
recommendations.

Step

Display / Result

^ Press the "Drying" 0 button until E02

appears in the display.

E02

^ Turn the program selector to "Intensive".

The programmed dosage
concentration appears in the
display, e.g.,

0.10 = 0.10 %

^ Press and hold the "Display" ß button

until the required value appears in the
display, e.g., 0.2.

0.20 The dosage concentration is
now set.

Dosage concentration setting:

Priming the dosage system

The dosage system does not need to be primed (cleared of air) unless:
– the dosage concentration is being set for the first time,
– the neutralizer container was inadvertently completely emptied .

Step

Display / Result

^ Press the "Drying" 0 button until E02

appears in the display.

E02

^ Turn the program selector to "Universal". d§3
^ Press the "Display" ß button.

The dosage system will be primed
automatically. Priming is complete
when

0 appears in the display.

^ Exit the programming mode.

Start "Rinse" program.

This will rinse away any agents that
entered the wash cabinet during
priming.