Thermo switch 59, Frequently asked questions, Thermo switch – Miele Laboratory Glassware Washer G 7883 User Manual

Page 59

Advertising
background image

Thermo switch

This machine has a resettable heater
limiter which will shut off the elements in
the event of over-heating. This could be
caused for example, if large articles
cover the heating elements or if the
filters in the wash cabinet are blocked.

If the following fault shows:

F01 - F03,

F18, F19, F27 (Water in the wash
cabinet is not heated, the program
cycle takes too long). Please proceed
as follows.

^ Remove the cause of the problem.
^ Remove the service plinth.
^ Press the reset button on the

temperature sensor on the right side.

If this switch trips repeatedly,
contact the Miele Technical Service
Department.

Frequently asked questions

59

Advertising