SECTION 5—OPERATING INSTRUCTIONS

Perfecto

2

Series

18

Part No 1143482

Connect Humidifier (If So Prescribed)

ƽ WARNING

For optimum performance, Invacare recommends that

each concentrator be on and running for a minimum of 30

minutes at a time. Shorter periods of operation may

reduce maximum product life.
DO NOT overfill humidifier.
DO NOT reverse the oxygen input and output

connections. Water from the humidifier bottle will travel

through the cannula back to the patient.

NOTE: For this procedure, refer to FIGURE 5.1, FIGURE 5.2 on
page 19, FIGURE 5.3 on page 20.

1. Remove cap from bottle.

2. Fill humidifier with distilled water to the level indicated

by the manufacturer. Replace the humidifier cap and
securely tighten.

FIGURE 5.1 Filling the Humidifier

3. Insert a flathead screwdriver in the plate groove on the top

edge of the filter access door and gently pry the filter
access door off (FIGURE 5.2).

4. Pull up and remove the humidifier bottle adapter

(FIGURE 5.2).

Humidifier

Bottle w/o Cap

Humidifier

Bottle with Cap