Initial startup of the concentrator, Note: for this procedure, refer to figure 5.6, Green light (o2) - normal operation – Invacare IRC5P User Manual

Page 23: Auto shut down, Use backup, Sensor feature)

Advertising
background image

SECTION 5—OPERATING INSTRUCTIONS

Part No 1143482

23

Perfecto

2

Series

Initial Startup of the Concentrator

NOTE: Concentrator may be used during the initial start warm-up
time (approximately 30 min.) while waiting for the O

2

purity to

reach maximum.

When the unit is turned on, the GREEN light will come on
(SYSTEM OK/O

2

greater than 85%). After five minutes, the

oxygen sensor will be operating normally and will control the
indicator lights depending on oxygen concentration values.

Explanation of Oxygen Purity Indicator
Lights - Model IRC5PO2

NOTE: For this procedure, refer to FIGURE 5.6.

GREEN light (O

2

) - Normal Operation.

YELLOW light (

) - Call supplier IMMEDIATELY. You may

continue to use the concentrator unless instructed otherwise
by your supplier. Be certain that backup oxygen is nearby.

RED light ( ) - Total unit shutdown. Switch IMMEDIATELY
to backup oxygen supply. Call supplier IMMEDIATELY.

GREEN light - with YELLOW light flashing - Call supplier
IMMEDIATELY. Oxygen sensor malfunctioning; you may
continue to use the concentrator.

FIGURE 5.6 SensO

2

Oxygen Purity Indicator (If Your Unit

Features the O

2

Sensor Feature)

O

2

Auto Shut Down

Use Backup

Below

Normal

Normal

RED

Call Supplier

YELLOW

GREEN

OXYGEN PURITY

Advertising
This manual is related to the following products: