PROCEDURE 3

FRONT RIGGINGS

3. Retighten the two (2) flat screws, washers and lock-

nuts.

NOTE: The settings for positioning the articulating
footplates on the half-clamps may vary for each footplate.

Footrest Support

Washers

Coved Washer

Locknut

FIGURE 3 - INSTALLING HINGE

Footplate Hinge

Socket Screw

Nylon

Adjustment

Screw

Footplate

Hinge

Half

Clamp

Articulating

Footplate

Flat Screws

Half Clamp

Locknuts

Washers

90

o

Footrest Support

Footplate Hinge

FIGURE 4 - INSTALLING FOOTPLATE/DEPTH

ADJUSTMENT/ANGLE ADJUSTMENT/

PERPENDICULAR AND/OR INVERSION/

EVERSION ADJUSTMENT

3. Flip the footplate hinge to the UP position.

NOTE: The footplate hinge will fall to the DOWN position.

4. Tighten the socket screw and locknut that secure the

footplate hinge to the footrest support until the footplate
hinge remains in the UP position.

5. Check the up and down motion of the footplate hinge to

make sure the user of the wheelchair can operate the
footplates easily.

NOTE: If the footplate's motion is too tight, loosen the socket
screw and locknut approximately 1/4-turn.

NOTE: If the footplate's motion is too loose, tighten socket
screw and locknut approximately 1/4-turn.

6. Install the adjustable angle flip-up footplate. Refer to the

following section of this procedure.

Installing Footplate (FIGURE 4)

1. Slide the half clamp over the footplate hinge.

2. Loosely tighten the two (2) flat screws that secure the

footplate to the half clamp.

3. Adjust the footplates to the necessary angle and depth

for the user. Refer to the following sections of this pro-
cedure.

Depth Adjustment (FIGURE 4)

1. Remove the two (2) flat screws, washers and lock-

nuts that secure articulating footplate to the footplate
hinge.

NOTE: Observe the angle of the articulating footplate for
reinstallation.

2. Move the articulating footplate to one (1) of four (4)

mounting positions.

NOTE: If desired depth is still not obtained, rotate the half
clamp on the footplate hinge 180

o

.

Angle Adjustment (FIGURES 4 AND 5)

1. Loosen, but do not remove the two (2) flat screws

and locknuts that secure footplate to the footplate
hinge.

2. Position the articulating footplate to the necessary

angle to accommodate the user.

3. Retighten the two (2) flat screws and locknuts.

Footplate

Footrest Support

Side View Of

Footplate

and Footrest

Support.

FIGURE 5 - ANGLE ADJUSTMENT

F

R

O

N
T

R

I

G
G

I

N

G

S