Specifications (continued) – Invacare PLATINUM 5 User Manual

Page 10

Advertising
background image

10

SPECIFICATIONS

SPECIFICATIONS (CONTINUED)

Label

Status

Indicator Lights (LED)

Symbol

I/O

SYSTEM OKAY

GREEN

SYSTEM FAILURE

RED

5LX INDICATORS:

Flow Range:

Low Flow Alarm:
Average Power
Consumption:
Pressure Relief
Mechanism Operational at:
Change in maximum

recommended flow

when back pressure
of 7kPa is applied:
Filters:

Safety System:

Width:
Height:
Depth:
Weight:
Shipping Weight:

0.5 - 5 L/min.

Flowrates less than 1 L/min., are not
recommended.
0 L/min. to 0.5 L/min., Rapid Audible
Alarm Beeping (No Accessories Connected)
395W

241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)

0.7 L/min.
Cabinet (2), Outlet HEPA and

Compressor Inlet.
Current overload or line surge shutdown.

High temperature compressor shutdown.

High Pressure Alarm w/compressor shutdown.

Low Pressure Alarm w/compressor shutdown.

Battery Free

Power Loss Alarm.

Sen

2

Oxygen System (SensO

2

Model)

Low Flow Alarm.
46.7 cm ± 1 cm (18 3/8” ± 3/8”)
67.0 cm ± 1 cm (26 3/8” ± 3/8”)
36.5 cm ± 1 cm (14 3/8” ± 3/8”)
23.2 kg ± 0.5 kg (51 lbs. ± 1 lb)
26.3 kg ± 0.5 kg (58 lbs. ± 1 lb)