Concentrator maintenance – Invacare PLATINUM 5 User Manual

Page 19

Advertising
background image

19

Cabinet Filter

(located on

both sides of

concentrator)

FIGURE 1 - CABINET

FILTER LOCATION

CONCENTRATOR MAINTENANCE

NOTE: Normal maintenance is to be performed by the user or caregiver.

The following outlines normal maintenance procedures.
NOTE: The Invacare Concentrators are specifically designed to minimize

routine preventive maintenance at intervals of once per year. In places with

high dust or soot levels, maintenance may need to be performed more

often. Preventive maintenance must be performed at a minimum of one (1)

year in service to assure years of additional reliability. Only authorized or

factory-trained personnel should perform preventive maintenance on the

concentrator. Power should be disconnected before beginning preventive

maintenance on the Invacare concentrator.

A.

CLEANING THE CABINET FILTERS

NOTE: There are two (2) cabinet

filters. One (1) is located on each

side of the cabinet (FIGURE 1).
1. Remove each filter and

CLEAN AT LEAST ONCE

A WEEK depending on

environmental conditions

(FIGURE 2).

NOTE: Environmental conditions that

may require more frequent cleaning of

the filters include but are not limited to:

high dust, smoking, air pollutants, etc.
2. Clean the cabinet filters with a vacuum cleaner or wash in warm

soapy water and rinse thoroughly (FIGURE 2).

3. Dry the filters thoroughly before reinstallation.

CAUTION

DO NOT operate the concentrator without the filters installed.

MAINTENANCE

FIGURE 2 - CLEANING THE CABINET FILTERS