21 troubleshooting – Invacare PLATINUM 5 User Manual

Page 21

Advertising
background image

21

TROUBLESHOOTING

1. Internal repairs required.

1. Call Service Provider.

GREEN LIGHT
with YELLOW
Light Flashing.

Alarm:
RAPID

Beep....Beep...
Beep...Beep

LOW FLOW ALARM
1a. Kinked or blocked

tubing, cannula or
humidifier.

1b. Flowmeter set at

0.5 L/min. or less.

1a. Inspect for kinks or

blockages. Correct,
clean or replace item.
Once corrected, turn
power OFF for 60
seconds and then turn
power back ON.

1b. Reset flowmeter to

prescribed flowrate.
Flowrates less than
1L/min. are not
recommended.

SYMPTOM:

YELLOW or RED
LIGHT
Illuminated.

Alarm:
CONTINUOUS
On red light only.

PROBABLE CAUSE:

1. Low oxygen purity.*
2. Kinked or blocked

tubing, cannula or
humidifier.*

3. Flowmeter set at

1/2 L/min. or less.*

4. Unit overheating

due to blocked air
intake.

5. Insufficient power at

outlet.

6. Internal repairs required.

SOLUTION:

1. Clean or replace filters.
2. Inspect for kinks or

blockages. Correct,
clean or replace item.
Once corrected, turn
power OFF for 60
seconds and then turn
power back ON.

3. Check flowmeter is set

to 1.0 L/min. or more.
Refer to Recommended
End-User Guidelines in
the Operating Instructions
of this manual.

4a. Remove and clean

cabinet filters.

4b. Move concentrator at

least 7.5 cm (3 in) from
walls, draperies, and
furniture.

5. Do not use extension

cords. Move to another
electrical outlet or circuit.

6. Call Service

Provider.

* Only applies to
SensO

2

models.