Fst - frame status information – SIGMA LBA-708 User Manual

Page 251

Advertising
background image

n..n

=

number of data bytes

(DAB) =

data byte - 8-bit data byte

Key Type Value

Description

FrameNumber I frame

number

-1 = gain frame
0 = reference frame
1 to n. ‘n’ is the size of the frame buffer.
default = current frame

Replace

B

default = false

true = replace
If the camera or the camera resolution

stored in the file do not match the current

configuration then the LBA-PC must replace

the current camera or change the

resolution before the frame can be loaded.
Either of these changes results in clearing

the frame buffer (i.e. all current frames are

discarded). If the pixel scale stored in the

file does not match the current pixel scale

then the pixel scale will be set to the value

in the file. If the

energy

calibration stored

in the file does not match the current

energy

calibration then the

energy

calibration will be set to the value in the

file.

false = do not replace
If the camera or camera resolution do not

match then the LBA-PC will issue an error

and the frame is not loaded. The pixel

scale and

energy

calibration values will not

be changed.

A.5.4.7 FST - frame status information

:FST <configuration>

1,3

:FST? [FrameNumber=f]

Returns:

Operator’s Manual

LBA-PC

Doc. No. 10654-001, Rev 4.10

251