Ieee 488.2 common commands, Lba-pc command and data formats – SIGMA LBA-708 User Manual

Page 180

Advertising
background image

DCL, GET, SDC, TCT

10.4.2 IEEE 488.2 Common Commands

The following IEEE 488.2 common commands are supported by the LBA-PC.

*IDN?

-

Identification

Query

*RST

-

Reset

*CLS -

Clear Status Registers

*ESE -

Event Status Enable Write

*ESE? -

Event Status Enable Query

*ESR? -

Event Status Register Query

*SRE -

Service Request Enable Write

*SRE? -

Service Request Enable Query

*STB? -

Status Byte Register Query


All other IEEE 488.2 common commands are not recognized by the LBA-PC and will return a

Command Error.

10.4.3 LBA-PC Command and Data Formats

The LBA-PC does not support the IEEE 488.2 specification of sending multiple commands separated

by semicolons. Each command must be sent separately and terminated by asserting the EOI line with

the last byte sent.
The following are some typical formats for command and response transmissions:
cmd:

*CCC(^END)

cmd:

*CCC?(^END)

cmd: *CCC _ key=value(^END)

cmd: *CCC _ key=value;key=value;…;key=value(^END)

rsp:

CCC _ key=value;key=value;…;key=value;;(^END)

rsp:

CCC _ FrameNumber=f;#dn..n (DAB)…(DAB)(^END)

Where:

*

=

Precedes a 488.2 common command

:

=

Precedes a LBA-PC command

CCC

=

Three character command code, case is ignored

?

=

Query character to illicit a response

Operator’s Manual

LBA-PC

180

_

=

ASCII SPACE character (0x20)

key

=

Command dependent parameter

Advertising
This manual is related to the following products: