Camera settings restrictions, Image synchronization considerations, Pyrocam i with digital lba-pc’s – SIGMA LBA-708 User Manual

Page 29: Pyrocam i setup requirements

Advertising
background image

2.6.1.3.5

Camera settings restrictions

Under no circumstances, make any changes to the Advanced. . . Camera settings for the
Pyrocam I.

2.6.1.4 Image synchronization considerations

The Pyrocam I’s CCIR video output is always producing video images at the rate of 25 frames per

second. Furthermore, it only changes output image after acquiring and processing a new image.

Thus the rate of a new output beam image is a function of many variables, including Pulse or
Chop rates, image processing time, and results computational times.
If you are using a Pulsed laser with your Pyrocam I, the Pyrocam will continue to output the

image of the last laser pulse that it receives. As a result, your LBA will continue to collect data

frames that are multiples of the last pulse received by the Pyrocam. To relieve this, you may

want to set up a Trigger input to the LBA. This will cause, at the most, only one image to be
acquired for each Laser Pulse. This can be a tricky situation since the Pyrocam does take a little

time to process each image, and that time will vary somewhat from image to image. Also, if your

laser is triggered faster than the Pyrocam can acquire and update its CCIR display, it will skip

pulses. Thus the LBA can still end up acquiring multiples of some of the processed pulses. You

can also try the solution described below for the Chopped-operating mode.
The exact same type of problem occurs in Chopped mode. The chop frequencies must be either
24Hz or 48Hz. Thus, the CCIR image can never be updating as fast as the chopper is triggering

the Pyrocam. Consequently, the CCIR output is always creating multiple frames of the same

beam profile. In this case, you may want to set the LBA’s Capture interval to a value that will

eliminate or reduce multiple frame captures.

2.6.2

Pyrocam I with Digital LBA-PC’s

LBA-500/7XXPC-D w/ Digital Camera Option

2.6.2.1 Pyrocam I setup requirements:

To use this interface method, your LBA-500/7XXPC must be equipped with the digital camera

option (-D), and your Pyrocam I must be of a later design that has digital output capability. Early
versions of the Pyrocam were not equipped with digital outputs. If your Pyrocam I does not have

a 50 pin connector on its rear cover it can not be interfaced as a digital camera. However, you

can still interface it by using the analog method described in the LBA-300PC interface topic.

Alternately, you can contact Spiricon’s service department and arrange to have your Pyrocam I

upgraded with digital camera output capability.

NOTICE: When interfacing with a Pyrocam I, DO NOT USE THE ULTRACAL! Feature of the LBA-PC. Rather, it
is ESSENTIAL that you periodically re-calibrate the Pyrocam I. Calibrating the Pyrocam I is the only calibration
that should be used with this type of interface.

Operator’s Manual

LBA-PC

Doc. No. 10654-001, Rev 4.10

29