SMC Networks Invivo HNC-63-INT User Manual

Page 17

Advertising
background image

17

Rev 05

Run an SNR CHECK scan using the parameters listed below.
TIP: Save this SNR CHECK scan for future QA scans.

Position slices for the SNR CHECK as shown on page 15.

Geometry
Coil Selection

3IB-MRIDevic

-channel Combination

2345

-connection

d

Homogeneity correction none
FOV (mm)

400.00

RFOV (%)

100.00

Foldover suppression

no

Matrix scan

256

-reconstruction

256

Scan percentage (%)

100.00

Stacks

1

-type

parallel

-slices

3

-slice thickness (mm)

5.00

-slice gap

user defined

- - gap (mm)

18.00

-slice orientation

sagittal

-foldover direction

AP

Slice scan order

default

PlanAlign

no

REST slabs

0

Patient position

head first

-orientation

supine

Contrast
Scan mode

MS

-technique

FFE

Contrast enhancement

no

Fast Imaging mode

none

Echoes

1

-partial echo

yes

TE

user defined

-(ms)

15.00

Flip angle (deg)

30.00

TR

user defined

-(ms)

200.00

Half Scan

no

Water fat shift

user defined

-(pixels)

2.000

Shim

no

SPIR

no

MTC

no

SAR mode

default

Gradient mode

regular

Motion
Cardiac synchronisation

no

Respiratory compensation

no

Flow compensation

yes

NSA

2

SMART

no

Dyn/ang
Angio

no

Quantitative flow

no

Manual start

no

Dynamic study

no

Postproc
Preparation phases

full

MIP/MPR

no

Images:

O:M no no no

Autoview image

M

Reference tissue

Grey matter

Preset window contrast

soft

Reconstruction mode

immediate

***Save raw data***

yes

Push to workstation

no

Hardcopy protocol

no

Ringing filter

no

Silent mode

no

Offc/ang
Stacks

1

Stack Offc.AP (P=+mm)

-20.00

RL (L=+mm)

0.00

FH (H=+mm)

0.00

Ang. AP (deg)

0.00

RL (deg)

-0.00

FH (deg)

-0.00

***IMPORTANT: Be certain to SAVE RAW DATA under Post Processing.***