Warning on power connection, Warning on power connection for medical use, For customers in taiwan only – Sony UP-D77MD User Manual

Page 3

Advertising
background image

3

susceptible equipment. Plug this unit and the
susceptible equipment into different branch circuit.
Consult your dealer. (According to standard
EN60601-1-2 and CISPR11, Class B, Group 1)

Caution
When you dispose of the unit or accessories, you must
obey the law in the relative area or country and the
regulation in the relative hospital.

Warning on power connection

Use a proper power cord for your local power supply.
1. Use the approved Power Cord (3-core mains lead) /

Appliance Connector / Plug with earthing-contacts
that conforms to the safety regulations of each
country if applicable.

2. Use the Power Cord (3-core mains lead) / Appliance

Connector /Plug conforming to the proper ratings
(Voltage, Ampere).

If you have questions on the use of the above Power
Cord/Appliance Connector/Plug, please consult a
qualified service personnel.

Warning on power connection for
medical use

Please use the following power supply cord.
With connectors (plug or female) and cord types other
than those indicated in this table, use the power supply
cord that is approved for use in your area.

*Note: Grounding reliability can only be achieved when the equip-
ment is connected to an equivalent receptacle marked ‘Hospital Only’
or ‘Hospital Grade’.

For Customers in Taiwan only

United States

Canada

Plug Type

HOSPITAL GRADE*

HOSPITAL GRADE*

Female end

E62405, E35708

LR53182, LL022442,
LL088408

Cord type

E159216, E35496
Min.Type SJT
Min.18 AWG

LL112007-1, LL20262,
LL32121, LL84494
Min.Type SJT
Min.18AWG

Minimum cord set
rating

10A/125V

10A/125V

Safety approval

UL Listed

CSA