Setup of the Unit

Perform the following settings when the unit is used for
the first time after being purchased, or when the IP
address assigned to the UP-D77MD is changed.

1

Open the DICOM Settings page.
For detailed information on how to open the page,
see “Opening the Setup Window” (on page 27).

2

Perform the required settings related to TCP/IP.

1

Enter the IP address of the UP-D77MD on the
network in the IP Address text box.

2

Enter the sub-net mask in the Sub-net Mask text
box.

3

Enter the IP address of the default gateway if a
gateway is used on the network.

3

Enter the DICOM AE title in the AE Title text box.
Up to 16 characters can be entered.

4

Click the [Save settings] button.
The settings are saved.

Setup of E-Mail

The E-Mail settings page allows you to set up E-mail
service so that an e-mail message will be sent
automatically when an error occurs in the UP-D77MD.

1

Open the E-Mail settings page.
For detailed information on how to open the page,
see “Opening the Setup Window” (on page 27).

2

Set criteria determining when an e-mail message
should be sent.

Place the check marks on the check box
corresponding to an error that requires notification
via an e-mail message.

IP Address text box

Sub-net Mask text box

Default Gateway text box

DICOM AE text box

Error item

Content

Motor/Sensor error

When an error occurs in the motor/
sensor system.

Film/Paper
Jamming

When a paper has jammed.

More than 30,000
printed with the
current thermal
head

When the total number of prints
with the current thermal head
exceeds 30,000.

Motor/sensor error check box

Film/Paper Jamming check box

More than 30,000 printed with the
current thermal head check box