Program Example 4:
IBM PC,Microsoft Basic,
via GPIB

This program requires the Capital Equipment
Corporation GPIB card for the IBM PC or XT. It
has firmware in ROM to interface high level
languages to the GPIB.

Subroutine calls in Microsoft BASIC are done to
memory locations specified by the name of the
subroutine. The address is relative to the segment
address specified by the DEF SEG statement
preceding CALL.

In this program, the CEC card's ROM starts at
OC0000H, the system controller's address is 21,
and the SR530 has been assigned as GPIB
address 23.

To monitor the GPIB activity with an RS232
terminal, SW1-6 should be down, and the ASCII
terminal should be attached to the rear panel
RS232 connector.

10

EXAMPLE PROGRAM TO READ THE SR510 OUTPUT AND RAMP THE X6 ANALOG OUTPUT

20

USING IBM PC BASICA AND THE CAPITAL EQUIPMENT CORP. GPIB INTERFACE CARD

30

40

50

ON THE SR510 REAR PANEL, SET SWITCHES #4 AND #6 ON SW1 TO DOWN (DEVICE

60

ADDRESS = 23, RS232 ECHO ON) AND SWITCH # 1 ON SW2 TO DOWN (RS232 BAUD

70

RATE = 9600). ALL OTHER SWITCHES SHOULD BE UP.

80

NOTE THAT THE RS232 ECHO IS FOR DEBUGGING AND DEMOSTRATION PURPOSES,

90

UNDER NORMAL CONDITIONING, SWITCH # 6 OF SW1 SHOULD BE UP SINCE THE RS232

100

ECHO SLOWS DOWN THE GPIB INTERFACE.

110

120 DEF SEG = &HC000

′BASE ADDRESS OF CEC CARD

130 INIT=0: TRANSMIT=3: RECV=6:

′ADDRESSES OF CEC FIRM WA

RE ROUTINES

140 ADDR%=21: SYS%=0

′CONTROLLER ADDRESS

150 INZ$ =

″IFC UNT UNL MTA LISTEN 23 DATA ′Z′ 13″

160

170 Q$ =

″IFC MTA LISTEN 23 DATA ′Q′ 13″

180 X6$ =

″IFC MTA LISTEN 23 DATA ′X6,″

190 LISN$ =

″IFC UNT UNL MLA TALK 23″

200

210

220 CALL INIT(ADDR%,SYS%)

′INIT X6 OUTPUT TO ZERO

230 CALL TRANSMIT(INZ$,STATUS%)

′RESET SR510

240 GOSUB 540

′CHECK TRANSMIT STATUS

250

260 X = 0

′INIT X6 OUTPUT TO ZERO

270

280 CALL TRANSMIT(Q$,STATUS%)

′READ OUTPUT

290 GOSUB 540
300 GOSUB 450

′GET RES

ULT

310 V1 = VAL(ANS$)

′INTO V1

320

330

340 PRINT

″OUTPUT = ″

;V1

350

360 X = X + .0025

′INCREMENT X6 OUTPUT BY 2.5 MV

370 IF X>10 THEN X 0

′RESET RAMP

380 X$ = X6$ + STR$(X) +

″′ 13″

′MAKE X6 COMMAND STRING