SR510 Specification Summary

General

Power

100, 120, 220, 240 VAC (50/60 Hz); 35 Watts Max

Mechanical

17" x 17" x 3.5" (Rack Mount Included) 12 lbs.

Warranty

Two years parts and labor.

Signal Channel

Inputs

Voltage:

Single-ended or True Differential

Current:

10

6

Volts/Amp

Impedance

Voltage:

100 M

+ 25 pF, ac coupled

Current:

1 k

to virtual ground

Full Scale

Voltage:

100 nV (10 nV on expand) to 500 mV

Sensitivity

Current:

100 fA to 0.5

µ

A

Maximum

Voltage:

100 VDC, 10 VAC damage threshold

Inputs

2 VAC peak-to-peak saturation

Current:

10 mA damage threshold

1

µ

A ac peak-to-peak saturation

Noise

Voltage:

7 nV/

Hz at 1 kHz

Current:

0.13 pA/

Hz at 1 kHz

Common Mode

Range:

1 Volt peak; Rejection: 100 dB dc to 1KHz

Above 1KHz the CMRR degrades by 6 dB/Octave

Gain Accuracy

1% (2 Hz to 100KHz)

Gain Stability

200 ppm/°C

Signal Filters

60 Hz notch, -50 dB (Q=10, adjustable from 45 to 65 Hz)

120 Hz notch, -50 dB (Q=10, adjustable from 100 to 130 Hz))

Tracking bandpass set to within 1% of ref freq (Q=5)

Dynamic Reserve

20 dB

LOW

(1

µ

V to 500 mV sensitivity)

40 dB

NORM (100 nV to 50 mV sensitivity)

60 dB

HIGH

(100 nV to 5 mV sensitivity)

Bandpass filter adds 20 dB to dynamic reserve

Line Notch filters increase dynamic reserve to 100 dB

Reference Channel

Frequency

0.5 Hz to 100 kHz

Input Impedance

1 M

, ac coupled

Trigger

SINE:

100 mV minimum, 1Vrms nominal

PULSE:

±1 Volt, 1

µ

sec minimum width

Mode

Fundamental (f) or 2nd Harmonic (2f)