Guidance and manufacturer’s declaration - electromagnetic immunity

The LMD-3250MD is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-
3250MD should assure that it is used in such as environment.

Immunity test

IEC 60601 test

level

Compliance level

Electromagnetic environment - guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2.5 GHz

3 Vrms

3 V/m

Portable and mobile RF communications equipment should be
used no closer to any part of the LMD-3250MD, including
cables, than the recommended separation distance calculated
from the equation appliance to the frequency of the transmitter.

Recommended separation distance

d = 1.2

P

d = 1.2

P 80 MHz to 800 MHz

d = 2.3

P 800 MHz to 2.5 GHz

Where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,

a

should be less than the compliance

level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with
following symbol:

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from

structures, objects and people.

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur

radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the LMD-3250MD is used exceeds the applicable RF compliance level above, the LMD-3250MD should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the LMD-3250MD.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.