Series 220/221 Instruction Manual

Chapter 2 Installation

Wiring Connections

Warning!

To avoid potential electric

shock, follow National Electric

Code safety practices or your

local code when wiring this

unit to a power source and to

peripheral devices. Failure to

do so could result in injury or

death. All wiring procedures

must be performed with the

power off.

The NEMA 4X enclosure contains an integral wiring compartment with one
dual strip terminal block (located in the smaller end of the enclosure). Two
3/4-inch female NPT conduit entries are available for separate power and
signal wiring. For all hazardous area installations, make sure to use an
agency-approved fitting at each conduit entry. If conduit seals are used, they
must be installed within 18 inches (457 mm) of the enclosure.

Input Power Connections

To access the wiring terminal blocks, locate and loosen the small set screw
which locks the small enclosure cover in place. Unscrew the cover to expose
the terminal block.

DC Power Wiring

Connect 4-20 mA loop power (12 to 36 VDC) to the +Pwr and –Pwr termi-
nals on the terminal block. Torque all connections to 4.43 to 5.31 in-lbs (0.5
to 0.6 Nm). The DC power wire size must be 20 to 10 AWG with the wire
stripped 1/2 inch (14 mm). The nominal voltage required to operate the 4-20
mA loop is 12 volts at the meter. The 4-20 mA loop is optically isolated from
the flow meter electronics.

Figure 2-14. DC Power Connections