Series 220/221 Instruction Manual

Chapter 3 Operation

Calibration Menu

Password

ENTER

Use
keys to access menus

Calibration

Menu

Meter Size

or Pipe ID

ENTER

Run Mode

Meter Factor

xxxx

Vortex Coef Ck

xx

Low Flow Cutoff

xx

Serial Number

xxxxxxxxx

Series 220 - meter size
Series 221 - pipe internal diameter (inches)

Meter calibration constant
Series 220 - pulses/ft

3

Series 221 - pulses/ft

Adaptive filter setting

Low Flow Cutoff
setting displayed
in volumetric flow
units (view only)

Process Temp(°F)

xxx

Used to correct Meter Factor for thermal
expansion of meter body for Series 220

Density

xxx

Units are lb/ft

3

Ref Density

xxx

Units are lb/ft

3

. Used for calculation of

standard cubic meters, standard cubic feet,
and normal cubic meters.

Viscosity

xxx

Units are centipoise (Cp)

< Vol (xxx/xxx) >

xxx

< Mass (xxx/xxx)

xxx

Low Flow Cutoff
setting displayed
in mass flow
units (view only)

The Calibration Menu contains the calibration coefficients for the flow meter. These
values should by changed only by properly trained personnel. The Vortex Coef Ck and
Low Flow Cutoff are set at the factory. Consult the factory for help with these settings if
the meter is showing erratic flow rate. The units of measurement used in the Calibration
Menu are preset and are as follows: Density = lbm/ft

3

, Reference Density = lbm/ft

3

,

Viscosity = centipoise.

IM-22

3-9