Appendix A Specifications

IM-22

A-3

Power Requirements

Loop powered, 12 to 36 VDC.


Output Signals

Analog: field rangeable linear, 4-20 mA output signal, 1000 ohms
maximum loop resistance, selected by user for volumetric flow rate
or mass flow rate.


Pulse: field rangeable volume/pulse output for totalization is a 50-
millisecond duration pulse operating a solid-state relay capable of
switching 40 VDC, 40 mA maximum.


Display

Alphanumeric 2 x 16 LCD digital display.

Six push buttons (up, down, right, left, enter, exit) operable either di-
rectly on the display panel or with a hand-held magnet through the
display glass of the explosion-proof enclosure.

Viewing at 90° mounting intervals.


Totalizer

Based on user-determined flow units, six significant figures in scien-
tific notation. Total stored in non-volatile memory.


Wetted Materials

Series 220 In-Line Flow Meter:

316L stainless steel standard.

C276 hastelloy or A105 carbon steel optional.

Series 221 Insertion Flow Meter:

316L stainless steel standard.

Teflon

®

packing gland below 500° F (205° C).

Graphite packing gland above 500° F (205° C).


Enclosure

NEMA 4X cast enclosure.


Electrical Ports

Two 3/4-inch female NPT ports.

Mounting Connections

Series 220: Wafer or 150, 300, 600 lb ANSI flange.

Series 221 Permanent installation: 2-inch MNPT; 150, 300, 600 lb
ANSI flange with compression fitting probe seal.

Series 221 Hot Tap

(1)

Installation: 2-inch MNPT; 150, 300, 600 lb

ANSI flange and optional retractor with packing gland probe seal.

Note: (1) Removable under line pressure.


Mounting Position

(1)

Series 220 In-Line Flow Meter: No effect.

Series 221 Insertion Flow Meter: Meter must be perpendicular
within ± 5° of the pipe centerline.

Note: (1) For liquid applications, pipeline must remain full at all
times.


Certifications Construction

Inspection ( ANSI/ASME B31.3).

Materials (NACE MR-01-75[90]).

CE and FM approved.

CSA, CENELEC approval pending.


FM

approvals:

Class I, Division 1, Groups B, C, & D, T6 at Tamb = 60°C

Class II/III, Division 1, Groups E, F, & G

IP66, NEMA 4X

















Advertising