Installation, Warning – Sanyo MPR-1411R User Manual

Page 14

Advertising
background image

- 51 -

13

INSTALLATION

1. Removing the packaging materials and tapes
Remove all transportation packaging materials and
tapes. Open the doors and ventilate the unit. If the
outside panels are dirty, clean them with a diluted
neutral dishwashing detergent. (Undiluted detergent
can damage the plastic components. For the dilution,
refer to the instruction of the detergent.) After the
cleaning with the diluted detergent, always wipe it off
with a wet cloth. Then wipe off the panels with a dry
cloth.

2. Adjusting the leveling foot
Extend the leveling feet by rotating them

counterclockwise until they make contact with the floor.
(Fig. 1) Ensure the unit is level.

3. Fixing the unit
Two fixtures are attached to the rear of the frame.
Fix the frame to the wall by attaching a rope or chain
between the wall and the fixtures.(Fig. 2)

4. Ground (earth)
The ground (earth) is for preventing the electric shock
in the case of the electrical insulation is somehow
degraded. Always ground the unit at the time of
installation.

WARNING

Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply outlet is
not grounded, it is necessary to install a ground by qualified engineers.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such
grounding may cause electric shock in the case of an incomplete circuit.

Leveling foot

Fig.

Fig.

Fixture