ENVIRONMENTAL CONDITIONS

This equipment is designed to be safe at least under the following conditions (based on the IEC 61010-1):

1. Indoor use;

2. Altitude up to 2000 m;

3. Ambient temperature 5

o

C to 40

o

C

4. Maximum relative humidity 80% for temperature up to 31

o

C decreasing linearly to 50% relative

humidity at 40

o

C;

5. Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage;

6. Other supply voltage fluctuations as stated by the manufacture;

7. Transient overvoltages according to Installation Categories (Overvoltage Categories) II; For mains
supply the minimum and normal category is II;

8. Pollution degree 2 in accordance with IEC 60664.