Dimension (mm):

Compatible actuators:

Actuators that are compatible with TD1:

LA31

LA27

Cable for LA27
The cable used together with LA27 has to ordered separately:

LA27
(0705816)

Cable for LA31
The cable used in combination with LA31 is not possible to choose in the respectable ordering examples for the actuator. To
avoid any confusing write X when stating the cable and specify the cable as 0705825.

LA31
(0705825)

LA31 with fast motor cannot be used in combination with TD1

Always choose LA27 and LA31 without brake otherwise there will it not be enough power in the 2 x 9 Volt batteries
for emergency lowering. In some cases where a standard HB is to run head, foot motor and the external actuator
simultaneous down it can happen that the actuator not will start if a brake is mounted. In the TD1 280 system with
memory and infrared the current cut-off down is not suffi cient to start up the extra actuator if a brake is chosen.