COMPACT 2

TM

15

Operation

English

®

The chart below shows the replenishment time to be set for the requested replen-
isher rate per m

2

-film surface. The standard setting is 10 sec. with 500 ml replen-

isher rate per m

2

film surface. The replenisher rate has to be adjusted depending

on film material, chemicals and film throughput.

Dimmer Function of Display

With the dimmer function the intensity of the control panel display can be changed.
The intensity can be reduced to avoid possible exposure of high-sensitive films.
Normal-sensitive films are not affected by the display.

Setting the intensity of the display:
1.

Switch processor on.

2.

Press “arrow down” button (3) for about 5 seconds. The intensity changes.
Two intensity-levels are available: bright and dark. During film-throughput this
function is not available.

Machine Cover Safety Disconnection

The machine cover may only be removed for service and maintenance purposes.
The processor cannot be started without the cover. In the event that the machine
cover is removed during film-throughput, the film transport will be stopped. On the
display the error message “E1” will be displayed. The error will be reset when the
machine cover has been replaced.

Replenishment time and replenishment rate relation

Replenisher rate

(ml/m

2

)

Replenisher time (s)

Replenisher rate

(ml per cycle)

200

4(4)

50

300

6(5)

75

400

8(7)

100

500

10(9)

125

600

13(11)

150

700

15(12)

175

800

17(14)

200

900

19(16)

225

1000

21(18)

250

Values in brackets for 60 Hz power connection.

Values when pump is set at 100% pump capacity.