CALIBRATION
The CALIBRATION function in this unit measures LCD panel characteristics from low to high brightness values
and internal monitor processing handles CALIBRATION.
CALIBRATION does not rely on image quality settings since internal signals are used for a calibration.
CALIBRATION in this unit is made at D65 color temperature and calibrations for other color temperatures are
results calculated from this value.

_ Equipment required for calibration

Konica Minolta CA-210 display color analyzer

Konica Minolta CA-PU12 or CA-PU15 standard measurement probe

RS-232C cable (male to male, straight)

• Connect the RS-232C terminal on this unit to the RS-232C terminal on the CA-210 display color analyzer using a

straight cable.

• Be sure to set the following calibration values on the CA-210 display color analyzer. The analyzer will not produce

a correct calibration unless it has been calibrated. For details on setup procedures, refer to the CA-210 User's
Guide.

• Turn on this unit and perform adequate aging (about 1 hour) before starting calibration.
• Turn off the lights in the room and make sure that no external light can enter the standard measurement probe

before starting the calibration. External light that enters the probe will prevent correct calibration of low brightness
characteristics.

• LCD panel characteristics and instrument error in the display color analyzer may sometimes result in small

differences in values after calibration.
In a fine tuning of the monitor, also set GAIN and BIAS for R and B in the COLOR TEMP. VAR mode. The GAIN
and BIAS values for G cannot be altered.

• Do not place 3D Polarized Eyewear between the analyzer and the monitor during calibration.

CA-210 side

(Straight)

BT-3DL2550P/E Side

Pin number

Signal

Pin number

Signal

1

CD

1

N.C.

2

RXD

2

TXD

3

TXD

3

RXD

4

DTR

4

DSR

5

GND

5

GND

6

DSR

6

DTR

7

RTS

7

CTS

8

CTS

8

RTS

9

GND

9

N.C.

Konica Minolta CA-210 display
color analyzer

This unit

Konica Minolta CA-PU12 or CA-PU15
standard measurement probe

RS-232C cable

Calibration

W

R

G

B

x

0.3091

0.6759

0.2029

0.1495

y

0.3429

0.3126

0.691

0.0587

Lv

195.3

47.56

135.2

13.21