Ratio imaging: multi-frame ratio – Roper Photometric User Manual

Page 22

Advertising
background image

16 Advanced Camera Operation Manual

Ratio Imaging: Multi-Frame Ratio
This script performs ratio imaging with the CCD37-10 CCD. Only three small subregions are needed, but to
get a good statistical base for the images, you must take a sequence of 100 identical exposures. The subimage
setup is:

+--------------------------------------------+

s1,p1

s2,p2

|

###

| <-- region 3

380,300

440,390

CCD37-10

|

###

|

Image

Array---->

|

#####

|

<--

region

2

175,26 320,290

|

#####

|

|

######

| <-- region 1

300,5

400,25

|

######

|

+--------------------------------------------+<-- origin point: 0,0

storage array

These coordinates are INCLUSIVE. 300,5 to 400,25 is a region of 101 serial columns and 21 parallel rows.
This experiment swings in a filter (relatively slowly), sends a trigger pulse, exposes for 100 milliseconds, reads
out, then repeats.

H/W A:

experimental hardware goes to first position, sends trigger pulse,

H/W B:

the first position runs for 100 milliseconds,

H/W C:

change filters (about 40 milliseconds of delay)

H/W D:

the second position runs for 100 milliseconds

H/W E:

change filters (about 40 milliseconds of delay), loop back to A

CCD A:

prepare for exposure, clear CCD, wait for trigger...

CCD B:

expose for 100 milliseconds

CCD C:

shift_image_to_storage

, readout, clear CCD, wait for trigger...

CCD D:

expose for 100 milliseconds

CCD E:

shift_image_to_storage

, readout, loop back to A

The only problem is that you cannot use a signal to differentiate between CCD B and CCD D. In the future,
camera hardware might be enhanced so that it produces a flash signal in CCD E. Then you would use the
enhancement to trigger step H/W A.

script_begin();
shutter_open();
loop_begin(200);

/* SEQUENCE OF 100 IMAGE PAIRS(200 total loops) */

clear_parallel(4);

/* CCD A&C: make sure CCD is cleared

*/

clear_until_trig();

/*

clear until the experiment is ready

*/

expose(100);

/* CCD B&D:

expose with this filter

*/

shift_image_to_storage();

/* takes 1.23 msec, shift image under mask

*/

shift_mode_s();

/* CCD C&E: READOUT...

*/

shift(37);

/*

move REGION 1 up to serial register

*/

pixel_readout(300,101,1, 21,1);

/*

s1:s2 of 300:400 is 101 total pixels

*/

pixel_display( 101, 21 );

/*

actual size of the region

*/

pixel_readout(175,146,1,265,1);

/*

REGION 2 is at the serial reg

*/

pixel_display( 146, 265 );
shift(9);

/*

move REGION 3 up to the serial regist*/

pixel_readout(380, 61,1,265,1);
pixel_display( 61, 265 );
loop_end();
shutter_close();
script_end(1);

/* leave CCD in continuous clear mode

*/

NOTE: The CCD37-10 has a 512x512 image array, but a 512x544 storage array. Discard (5+32) rows to move
region 1 to the serial register.