SULLIVAN Comfort Operating Manual ––––––––––––––––––––––––––––––––––––––––––––––

22

Troubleshooting

If you have any problems with your SULLIVAN

®

Comfort™, try the following

suggestions. If it is still not working properly, consult your equipment supplier or
take it to an authorized service agent. Do not attempt to open the flow generator.

11

Problem

Possible cause

Solution

Flow generator does
not start when you
press START or delay
timer button
(and no indicator
lights are showing on
the front panel)

Nasal cushion not properly
fitted onto mask frame

Mask and headgear straps not
positioned correctly

Plugs missing from access
ports of mask frame

Nasal cushion is torn or
distorted

Air leaking around
nose and eyes

Delay timer is in use

Air filter strips are dirty

Air tubing is kinked or
punctured

Air tubing is not connected
properly

Plugs are missing from access
ports

Insufficient air
delivered from flow
generator

Wait for air pressure to build up gradually,
according to selected delay period. Or
press START to override delay period.
(See section 5, Daily operation.)

Replace air filter strips
(see section 9, Cleaning and maintenance)

Straighten or replace tubing

Tighten air tubing connections at both ends

Replace plugs

Fit cushion onto mask frame, aligning the
two pins on the cushion with the slots on
the mask frame
(see section 3, Preparing for use)

Adjust position of mask and headgear.
Note that if you are using the Bubble
Cushion, the flow generator needs to be
running before the cushion can be
properly fitted. (See section 4, Fitting the
mask
and section 5, Daily operation.)

Replace plugs

Replace cushion

Power cord not connected properly

Flow generator not switched on

Power outlet may be faulty

Connect power cord firmly at both ends

Switch main power switch at rear of flow
generator to on (I)

Try another power outlet