Displays error message:

SYSTEM ERRxxxx Call Service!

or Exxxx (where

xxxx defines an error)

Displays error message:

Check tube!! Key if done

The following message is displayed on the LCD after you try to update settings

or copy data to the Data Card:

Card Error Remove Card

The following message is displayed on the LCD after you try to update settings

or copy data to the Data Card:

Settings Error Remove Card

The following message is NOT displayed on the LCD after you try to update the

settings using the Data Card: Settings Success Remove Card

The following message is displayed on the LCD after you try to update settings

or copy data to the Data Card:

Settings Invalid Remove Card

Problem/Possible Cause

Solution

Component failure.

Record error number and call service.

Air tubing is loose or blocked.

Check that the air tubing is connected securely to your
mask and the air outlet to the front of the S8 Series II.
When you have checked the air tubing, press the Start/
Stop key to restart the device. If this does not clear the
message, disconnect the power cord and then
reconnect it to restart the device.

Data Card is not inserted
correctly.

Check that the patient is inserting the Data Card into the
Data Card module with

the arrow facing up, while the S8

Escape II is switched on and not delivering therapy
(standby mode).

The Data Card is only partly
inserted.

Ensure that the Data Card is inserted into the Data Card
slot as far as it will go.

You may have removed the
Data Card before settings
were copied to the
S8 Escape II.

Reinsert the Data Card and wait for the Settings

Success Remove Card

message to appear on the

LCD.

The card is not a ResScan Data
Card.

Only use a ResScan Data Card.

There is a data error on the
Data Card.

Contact your clinician immediately.

The settings were not
updated.

Contact your clinician immediately.

The identification details on
the Data Card do not match
the identification details on
your device.

Contact your clinician immediately.