STATUS Indicator

POWER Indicator

Action to Take

STATUS

Lit red

POWER

When power is on:

Lit green

When power is off:

Lit amber

It is timer to replace the lamp.
The lamp timer has exceeded 1000 hours.
Replace the lamp cartridge.

The lamp’s service life has ended.
The lamp timer has exceeded 1100 hours. The lamp
will not light. Replace the lamp cartridge.

STATUS

POWER

Lit red

Lit amber after 90 seconds

of blinking in amber

When the STATUS Indicator is Lit/Blinks

STATUS

Blinks red

(0.5 s lit / 0.5 s off)

POWER

Lit amber

The fan is stopped.
Unplug the power cable and request service from
Olympus Digital Technical Support.

STATUS

Blinks red

(1 s lit / 1 s off)

POWER

Lit red

The lamp cover is open.
The lamp cover is not attached properly.
Attach it properly.