Omega Speaker Systems FMG3000 User Manual

Page 7

Advertising
background image

7

6. General Installation and Grounding Tips

Sensor conditioning
The Magmeter output signal may be unstable immediately after installation. Allowing the sensor to soak in a full pipe (or in any
container of water) for 24 hours will stabilize the performance.

Very low conductivity fl uids may require a longer conditioning period. The Magmeter may not operate properly in fl uids where the
conductivity is less than 20 μS/cm.

Grounding
The FMG-3000 Magmeter is unaffected by moderate levels of electrical noise. However, in some applications it may be necessary to
ground portions of the system to eliminate electrical interference. The grounding requirements will vary with each installation.

7. Wiring the FMG-3002, 3102 and 3202 Magmeter with 4 to 20 mA Loop Output

The FMG-3002, 3102 and 3202 Magmeters are traditional 2-wire passive 4 to 20 mA loop transmitters.

External loop power (24 VDC ±10%) is required.

The maximum loop resistance the Magmeter can accomodate is 300 Ω.

All FMG-3002, 3102 and 3202 Magmeters are shipped from the factory with the 4 to 20 mA output scaled for 0 to 5 m/s (0 to
16.4 ft/s). The Calibration charts on pages 8-12 list the 20 mA setpoint for each installation fi tting. Use this information to program
the 4 to 20 mA range of the loop device (PLC, Datalogger, recorder, etc.)

1

2

3

4

Loop + (24 VDC)

Loop - (Ground)

4 to 20 mA Loop

monitor

(Maximum 300 Ω)

FMG-3000 Series Magmeter

+

+ -

-

24 VDC ± 10%

Grounding rings on plastic pipe

(Install between flanges)

or

metal straps on metal pipe

Earth ground

Sensor
Grounding ring

(10 cm to 1.3 m)

(10 cm to 1.3 m)

4 in. to 50 in.

4 in. to 50 in.

Instrument

4.

1.

2.

2.

Do not terminate
shield at Magmeter

3.

Connect an additional wire (minimum AWG 14/1.5 mm

2

) from the

remote instrument ground to the Magmeter ground terminal.

Install fl uid grounding devices immediately upstream and

downstream of the Magmeter.

Connect

the

fl uid grounds to the Earth ground terminal on the

FMG-3000.

Use

fl anged grounding rings or metal electrodes on plastic pipes,

or metal clamps on metal pipes.

Fluid grounds must be in direct contact with the fl uid, and as
near to the Magmeter as possible.

The shield from the output cable must be terminated at the

remote instrument ONLY.

This shield must not be connected at both ends!

One or more of the following steps may be
applied if the FMG-3000 Magmeter is affected by
electrical noise:

The ground terminal on the outside of the

yellow housing is connected internally
to the grounding ring at the tip of the
sensor. Connect a wire (14 AWG/1.5 mm

2

recommended) from this terminal directly to
a local Earth ground.

Advertising